Hi - Thanks for getting back to me so quickly.
I was told that when the Leads are connected, the machine is reading that they're not connected.
I double checked that there's no obstruction on the leads and that they're placed in the correct position.
Unsure if this may be a wire replacement that may be needed or not
Model: Burdick - ELI 280
Model Type: EKG
Custom Field: AM12 (S/N:123280001247)
Notes:
Asset ID: 13116139
Site: Oak Street Health - Oak Street Health - Strawberry Mansion, PA
Department: Wilmington Depot - Oak Street hot swapDepreciation Start Date:Physical Location: EMSARDepreciation Term: 7yrsSerial #: 123080000940
Cleaned device externals
Cleared stored ePHI
Visual Inspection; Passed
AM12 Condition; Good (S/N:_____123280001247_______)
Battery Date Code = ___2025____
Battery Test - No Load - (> 12.5VDC) = ___12.86____VDC
Battery Test - 10Ω25W - (>11.7VDC) after 10 sec = ___12.54____VDC
Current Test - Power OFF - (< 250uA) = ___101.62____uA
Current Test - Power ON - (< 1.1A) = ____0.57___A
Current Test - Charging voltage (13.0 - 14.7 VDC) = ___13.60____VDC
Touchscreen Calibrated
Date/Time; Updated
Auto Test; Passed
Writer Test; Passed
Printhead in good condition, cleaned, Que verified
Performance Tested EKG Leads @ 30, 60, 80, 120 BPM; Passed
Lead-off Warning; Verified
Noise Test; Passed
Electrical Safety Test; Passed
Enclosure Leakage Current measured >300µA; Passed
Patient Leakage Current measured >10µA; Passed
PM Completed
🔍Technician Findings
Was not able to duplicate issue. AM12 and lead wires are working correctly with no issues. Unit will need new clips to complete Hotswap.
⚠️Problem Description
When the Leads are connected, the machine is reading that they're not connected.
I double checked that there's no obstruction on the leads and that they're placed in the correct position.
Burdick - ELI 280
Asset ID: 13116139
Serial #: 123080000940 https://www.ebiotrack.com/workorder.php?wo=72702
📄Description
ECG, BURDICK 280 INTERP 200 ECG STORAGE AM12/USB/LAN
Oak Street Health’s Consigned Inventory
Timeline
📞
Case Linked Feb 16, 12:14 PM
Case Number:00568714
Subject:EKG Leads Off
Status:Closed
🚨27.0 business days
📋
▶
2
Work Order Created Mar 25, 3:26 PM
Work Order Created → Status: Assigned
👷
Status: Assigned by Samuel Klein
From:Tech On Site
To:Assigned
Duration in Previous:0m
🔧
Status: Tech On Site by Samuel KleinMar 25, 4:47 PM
From:Assigned
To:Tech On Site
Duration in Previous:1h 21m
🔩
4x Parts Added Mar 25, 4:58 PM
Type:Parts, Labor
Status:Open
Qty:1
✔️
Status: Completed by Samuel KleinMar 26, 8:31 AM
From:Tech On Site
To:Completed
Duration in Previous:15h 44m
🔩
Parts Added Mar 26, 8:36 AM
Type:Parts
Status:Open
Qty:1
🏁
Status: Closed by Samuel KleinMar 26, 8:54 AM
From:Completed
To:Closed
Duration in Previous:22m
📄
Status: Ready to Bill by Diane PattonMar 26, 5:17 PM
From:Closed
To:Ready to Bill
Duration in Previous:8h 22m
⏳4.8 business days
📨
Status: Invoice Pending by Diane PattonApr 1, 5:31 PM
From:Ready to Bill
To:Invoice Pending
Duration in Previous:6d 0h
💰
▶
2
Status: Invoiced Apr 1, 6:11 PM
Status: Invoiced → Invoice Created
🧾
Invoice Created
Invoice Number:SM-258111
Bottlenecks
Tech On Site
Duration: 1.4 business days (threshold: 1.0 business days)
Exceeded by 3.2 business hours
Ready to Bill
Duration: 4.3 business days (threshold: 1.0 business days)