Q-00031132
Depot Repair
Follow-up to WO-01299306
Asset ID 12982598
Model: Invacare - PERFECTO 2 (A)
Notes: Failed inspection Unit failed last inspection and has not been picked up for repair yet. Unit hours are 25932 test at 5 liters per minute Output is. 83.7%
1.25 x hr labor
1x- SBD0358
1x - FIL2036
https://www.ebiotrack.com/omnicor/workorder.php?wo=7021
4/14 TD- Located correct w/o for sieve beds received as previous w/o was closed & invoiced. Unit is now repaired & good to deliver back to site.
4/2 TD- Chattered due to w/o being closed/invoiced. Unit is NOT repaired. Just received the parts delivery. Unit is still at SYR depot.
3/25 TD- Packaged sieve beds & will drop off at UPS today.
3/20 TD- Placed part order.
3/13 TD- Chattered Ryan to have part order created for sieve bed repour & labels.
🔧Work Performed
Removed & cleaned 4 way valve. Unit running but has low O2. INSTALLED REPOURED SIEVE BEDS.
UNIT IS NOW RUNNING AT 95% AT 5LPM
🔍Technician Findings
Bad sieve beds & dirty filter. 4 way valve sticking.
Unit has 25932hrs
⚠️Problem Description
Q-00031132
Depot Repair
Follow-up to WO-01299306
Asset ID 12982598
Model: Invacare - PERFECTO 2 (A)
Notes: Failed inspection Unit failed last inspection and has not been picked up for repair yet. Unit hours are 25932 test at 5 liters per minute Output is. 83.7%
1.25 x hr labor
1x- SBD0358
1x - FIL2036
https://www.ebiotrack.com/omnicor/workorder.php?wo=7021