FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
03/11/2026 SRobinson- Called Noah Kane, 5712654714. Spoke with Noah and he confirmed for 03/16 for Solomon. Dispatching WO and sending a confirmed email back.
02/06/2026 RFosbury working to reschedule.
02/04/2026 Brod - received email from Joshua stating the point of contact will be Noah Kane, Noah.kane@HealthTrustPG.com, 5712654714. offered to have a FSE onsite for 02/09
🔧Work Performed
Template Name: FMI32097_Code 1M_ Screw_Disen Template Description: FMI32097_ Code 1M_ Screw_Disen Problem Found: FMI32097_Heater door inspection. Action Taken: FMI 32097 completed. Inspected the Heater doors and found shoulder screw disengaged. Screw tightened per instructions. Verification Test: 1. Verified the canopy moves smoothly. 2. Verified transition mode message is not present. 3. Verified doors close smoothly. 4. Verified corner pins are aligned. 5. Verified no alarms. The device has passed all tests and is ready for customer use.
🔍Technician Findings
I observed the left screw under the heater door slightly loose and I torqued it back to proper specification. The door movement was not impacted at all and there weren't any other loose screws. The machine is fully operational.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📋
Outbound to +15712654714
Completed Mar 11, 2026
03/11/2026 SRobinson- Called Noah Kane, 5712654714. Spoke with Noah and he confirmed for 03/16 for Solomon. Dispatching WO and sending a confirmed email back.
Work Order: WO-01326629
Account: GE Healthcare FMI
Customer Case Number: 1-556712134181
Location: RESTON HOSPITAL CENTER
Address: 1850 TOWN CENTER PKWY, RESTON, VA 20190
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB22490016SA
Contact Name: Noah Kane
Contact Phone: 57
JFIF
TAB22490016SA_Screw 4_18MAR2026
Mar 18, 2026
JFIF
TAB22490016SA_Screw 2_18MAR2026
Mar 18, 2026
JFIF
TAB22490016SA_Screw 3_18MAR2026
Mar 18, 2026
JFIF
TAB22490016SA_Screw 1 (was loose but tighted back)_18MAR2026
Mar 18, 2026
JFIF
TAB22490016SA_Product Label Pic_18MAR2026
Mar 18, 2026
Timeline
📞
Case Linked Dec 15, 2:02 PM
Case Number:00551763
Subject:GE FMI 32097 - Reston
Status:Closed
📋
Work Order Created Dec 15, 2:29 PM
Work Order:WO-01326629
Type:FCO
Priority:PC
🚨32.0 business days
✅
Status: Ready for Release by Mandy NewmanJan 31, 12:06 PM