FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
03/11/2026 SRobinson- Called Noah Kane, 5712654714. Spoke with Noah and he confirmed for 03/16 for Solomon. Dispatching WO and sending a confirmed email back.
02/06/2026 RFosbury working to reschedule.
02/04/2026 Brod - received email from Joshua stating the point of contact will be Noah Kane, Noah.kane@HealthTrustPG.com, 5712654714. offered to have a FSE onsite for 02/09
02/03/2026 Brod - called 5404507350 to speak to Joshua Pereira went to vm left message to have FSE onsite for 02/04. following up in email to joshua.pereira@gehealthcare.com & christopher.roberts3@hcahealthcare.com
🔧Work Performed
Template Name: FMI32097_Code 1_inspect. Template Description: FMI32097_Code 1_inspect. Problem Found: FMI32097_Heater door inspection. Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification. Verification Test: 1. Verified the canopy moves smoothly. 2. Verified transition mode message is not present. 3. Verified doors close smoothly. 4. Verified corner pins are aligned. 5. Verified no alarms. The device has passed all tests and is ready for customer use
🔍Technician Findings
No problems found on the machine per my observation. The screws were torqued properly and the machine doors, as well as the machine, works as it should.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📞
Live Call_ Left Message
Completed Feb 3, 2026
02/03/2026 Brod - called 5404507350 to speak to Joshua Pereira went to vm left message to have FSE onsite for 02/04. following up in email to joshua.pereira@gehealthcare.com & christopher.roberts3@hcahealthcare.com
Work Order: WO-01326632
Account: GE Healthcare FMI
Customer Case Number: 1-556712134263
Location: RESTON HOSPITAL CENTER
Address: 1850 TOWN CENTER PKWY, RESTON, VA 20190
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24050024SA
Contact Name: Noah Kane
Contact Phone: (5
JFIF
TAB24050024SA_Screw 1_18MAR2026
Mar 18, 2026
JFIF
TAB24050024SA_Screw 3_18MAR2026
Mar 18, 2026
JFIF
TAB24050024SA_Product Label Pic_18MAR2026
Mar 18, 2026
JFIF
TAB24050024SA_Screw 2_18MAR2026
Mar 18, 2026
JFIF
TAB24050024SA_Screw 4_18MAR2026
Mar 18, 2026
Timeline
📞
Case Linked Dec 15, 2:02 PM
Case Number:00551763
Subject:GE FMI 32097 - Reston
Status:Closed
📋
Work Order Created Dec 15, 2:31 PM
Work Order:WO-01326632
Type:FCO
Priority:PC
🚨32.0 business days
✅
Status: Ready for Release by Mandy NewmanJan 31, 12:06 PM