FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
03/26/2026 Brod - called 4705325660 to speak to Laura Thum went to vm left message to have FSE onsite for the 03/30. following up in email laura.thum@northside.com
12/23/2025 EMartin called Laura Thum / 470.532.5660 confirmed 40 devices onsite confirmed onsite to start devices 01/31/25 sent confirmation email to laura.thum@northside.com
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📞
Live Call_ Left Message
Completed Mar 26, 2026
03/26/2026 Brod - called 4705325660 to speak to Laura Thum went to vm left message to have FSE onsite for the 03/30. following up in email laura.thum@northside.com
Work Order: WO-01327219
Account: GE Healthcare FMI
Customer Case Number: 1-556774717630
Location: NORTHSIDE HOSPITAL
Address: 1000 JOHNSON FERRY RD NE,, ATLANTA,, GA 30342
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB23090040SA
Contact Name: Laura Thum
Contact Phone: