FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Primary Delay12.7 days stuck in "Ready for Release"
⏱️
Total Delay23.6 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
14.3d / 1.0d SLA
Ready for Release
12.7d
Scheduled
5.8d / 2.0d SLA
Tech On Site
0.0d / 1.0d SLA
Completed
1.3d / 1.0d SLA
Closed
2.3d / 1.0d SLA
Ready to Bill
7.9d / 1.0d SLA
WO Correction Needed
0.1d
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
12/11/2025 EMartin sent email to Justin advising FSE onsite Tues 12/16/25 to complete devices
12/09/2025 MHamersley - recd email from Justin, confirming TABA71546, TAB23440046SA, TAB24220044SA & TAB24220041SA at his location. The other SNs are not in inventory. ... 12/09/2025 MHamersley - called 404-605-2708, call went to Justin's VM/LM; sent email: justin.michaels@piedmont.org ... 11/05/2025 MKlose called 404-587-6218; Willie call went to VM ... 11/03/2025 MKlose called 404-587-6218; Willie cell phone. He will confirm devices by Wednesday or will call cell again.
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed
📞Call Description
FMI 32097
📧
Email_Response Confirmed
Completed Dec 9, 2025
12/09/2025 MHamersley - recd email from Justin, confirming TABA71546, TAB23440046SA, TAB24220044SA & TAB24220041SA at his location. The other SNs are not in inventory.
📧
Email: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - PIEDMONT HOSPITAL INC - ATLANTA, GA - WO-01312649
Completed Dec 9, 2025
To: justin.michaels@piedmont.org
CC: central.scheduling@emsar.com; erin.martin@emsar.com; mary.hamersley@emsar.com
BCC:
Attachment: FMI32097 MIC UMDC English Customer Letter.pdf
Subject: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - PIEDMONT HOSPITAL INC - ATLANTA, GA - WO-01312649
Body:
Hi Justin,
I am contacting you on behalf of GE Healthcare about a Field Safety Notice issued for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices. The in
📧
Live Call_Followed Up With Email
Completed Dec 9, 2025
12/09/2025 MHamersley - called 404-605-2708, call went to Justin's VM/LM; sent email: justin.michaels@piedmont.org