FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
12/11/2025 EMartin sent email to Justin advising FSE onsite Tues 12/16/25 to complete devices
12/09/2025 MHamersley - recd email from Justin, confirming TABA71546, TAB23440046SA, TAB24220044SA & TAB24220041SA at his location. The other SNs are not in inventory. ... 12/09/2025 MHamersley - called 404-605-2708, call went to Justin's VM/LM; sent email: justin.michaels@piedmont.org ... 11/05/2025 MKlose called 404-587-6218; Willie call went to VM ... 11/03/2025 MKlose called 404-587-6218; Willie cell phone. He will confirm devices by Wednesday or will call cell again.
🔧Work Performed
FMI 78101 Not Required –
Device Location is Unknown
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.