Request from Taylor Duda: Repair WOs needed for the following:
Equipment: BBraun Vista Basic Infusion Pumps
1- #13073649 - screen light out
2- #13143557 - malfunction beep + screen
3- #13143552 - downstream occlusion
4- #13082988 - no note
5- #13073650 - does not hold charge.
will need 1 battery, 34502556 OEM only!
11/17 KSmith -- Melissa wants Taylor to complete these repairs whenever her next available is.
🔧Work Performed
Note on unit states " downstream occlusion". Unit does not appear to have physical damage only access grease on pump housing. Battery is at full 3hr30min. Ran unit multiple times and could not duplicate downstream occlusion error unless manipulated. No other faults were found. Infusion pump can be returned to service.
🔍Technician Findings
Note on unit states " downstream occlusion". Unit does not appear to have physical damage only access grease on pump housing. Battery is at full 3hr30min.
⚠️Problem Description
Request from Taylor Duda: Repair WOs needed for the following:
Equipment: BBraun Vista Basic Infusion Pumps
Request from Taylor Duda: Repair WOs needed for the following: Equipment: BBraun Vista Basic Infusion Pumps 3- #13143552 - downstream occlusion