Request from Taylor Duda: Repair WOs needed for the following:
Equipment: BBraun Vista Basic Infusion Pumps
1- #13073649 - screen light out
2- #13143557 - malfunction beep + screen
3- #13143552 - downstream occlusion
4- #13082988 - no note
5- #13073650 - does not hold charge.
will need 1 battery, 34502556 OEM only!
11/17 KSmith -- Melissa wants Taylor to complete these repairs whenever her next available is.
🔧Work Performed
No note on unit. Battery is at full 3hr30min. Infusion pump asset # 13082988 does not appear to have any physical damage only access grease on pump housing. Ran unit multiple times with no faults or errors found. Infusion pump can be returned to service.
🔍Technician Findings
No note on unit. Battery is at full 3hr30min. Infusion pump asset # 13082988 does not appear to have any physical damage only access grease on pump housing.
⚠️Problem Description
Request from Taylor Duda: Repair WOs needed for the following: Equipment: BBraun Vista Basic Infusion Pumps 4- #13082988 - no note