FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Primary Delay39.3 days stuck in "Ready for Scheduling"
⏱️
Total Delay43.4 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
39.3d / 1.0d SLA
Scheduled
0.8d / 2.0d SLA
Tech On Site
1.0d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
2.0d / 1.0d SLA
Ready to Bill
5.5d / 1.0d SLA
Billing Review
2.5d / 2.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
11/25/2025 MHamersley - recd email from Dave Bell, "I can confirm these are still in inventory. I also have TAB24340042SA. It has had this door repaired twice since it was put in service last year" ... 11/25/2025 MHamersley - called 615-879-5201, spoke to Dave, POC, sent email: dave.bell@trimedx.com
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01316795 - FCO - Giraffe OmniBed Carestation - NASHVILLE - ASCENSION SAINT THOMAS HOSPITAL MIDTOWN
Jan 21, 2026 7:00pm – 7:00pm
Work Order: WO-01316795
Account: GE Healthcare FMI
Customer Case Number: 1-556561930803
Location: ASCENSION SAINT THOMAS HOSPITAL MIDTOWN
Address: 2000 CHURCH ST, NASHVILLE, TN 37236
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TABB70538
Contact Name: Dave Bell
Contact