FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Primary Delay46.8 days stuck in "Ready for Release"
⏱️
Total Delay57.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
45.3d / 1.0d SLA
Ready for Release
46.8d
Scheduled
1.5d / 2.0d SLA
Tech Off Site
0.1d / 2.0d SLA
Completed
0.0d / 1.0d SLA
Closed
4.4d / 1.0d SLA
Ready to Bill
14.7d / 1.0d SLA
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
04/03/2026 Brod - sent email to Ryan to have FSE onsite for 04/06
02/18/2026 MHamersley - recd email from Ryan Pumphrey, confirming and ready to schedule; called 817-250-3185, call went to Richard Vasquez, Bio Med Mgr, VM/LM; sent f/u emails: ryanpumphrey@texashealth.org & richardvasquez@texashealth.org ... 12/16/2025 MHamersley - called 817-250-3185, Adrian Rueda, call went to Richard Vasquez, Bio Med Mgr, VM/LM; called 817-250-3993, Ryan Pumphrey, call went to VM/LM; sent f/u emails: richardvasquez@texashealth.org & ryanpumphrey@texashealth.org ... 12/10/2025 MHamersley - called 817-250-3993, Ryan Pumphrey, call went to VM/LM; called 817-250-3185, Adrian Rueda, call went to VM/LM; sent emails: ryanpumphrey@texashealth.org & adrianrueda@texashealth.org ... Secondary site POC: Ryan Pumphrey T: Biomed Supervisor, Fort Worth D: (817) 250-3993 CCL: (817) 444-8600 E: ryanpumphrey@texashealth.org
🔧Work Performed
FE to select the completion Code “1M”. Enter the following content into the SR fields
as shown, which are required for field action closure:
Template Name: FMI32097_Code 1M_ Screw_Disen
Template Description: FMI32097_ Code 1M_ Screw_Disen
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and found
shoulder screw disengaged. Screw tightened per instructions.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
Rendered PDF File Page 6 of 20 DOC3055845, Rev:2Released
🔍Technician Findings
FE to select the completion Code “1M”. Enter the following content into the SR fields
as shown, which are required for field action closure:
Template Name: FMI32097_Code 1M_ Screw_Disen
Template Description: FMI32097_ Code 1M_ Screw_Disen
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and found
shoulder screw disengaged. Screw tightened per instructions.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
Rendered PDF File Page 6 of 20 DOC3055845, Rev:2Released
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01318287 - FCO - DD: 4/27/2026 - - Giraffe OmniBed Carestation - FORT WORTH - TEXAS HEALTH HARRIS METHODIST HOSPITAL FORT WORTH
Apr 13, 2026 8:00am – 5:00pm
Work Order: WO-01318287
Account: GE Healthcare FMI
Customer Case Number: 1-556592452843
Location: TEXAS HEALTH HARRIS METHODIST HOSPITAL FORT WORTH
Address: 1301 PENNSYLVANIA AVE, FORT WORTH, TX 76104
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24500139SA
Contact Na
MSG
TAB24500139SA-1
Apr 14, 2026
MSG
TAB24500139SA-3
Apr 14, 2026
MSG
TAB24500139SA-2
Apr 14, 2026
PDF
Completion Report-TAB24500139SA
Apr 14, 2026
Timeline
📞
Case Linked Dec 1, 10:08 AM
Case Number:00547582
Subject:GE FMI 32097 - creating IPs
Status:Closed
⏳2.0 business days
📋
▶
2
Work Order Created Dec 2, 2:19 PM
Work Order Created → Status: Ready for Scheduling
📅
Status: Ready for Scheduling by Kimberly Hardison
From:Entered
To:Ready for Scheduling
Duration in Previous:0m
⚠️7.0 business days
✅
Status: Ready for Release by Erin MartinDec 10, 12:36 PM
From:Ready for Scheduling
To:Ready for Release
Duration in Previous:7d 22h
🚨47.0 business days
📅
Status: Ready for Scheduling by Mary HamersleyFeb 18, 1:05 PM