Hey there,
One of our vital sign machines won't read O2- we have swapped out several cords and it won't read. Can we get a loaner and send this out for repair? Thanks.
SN: SR723010125SA
GE B125M
Elaine McEntee I Lead Clinician
Sono Bello - Portland
5100 SW Macadam Ave Suite 180
Portland, OR 97239
direct 503.575.9515 I fax 503.295.5823
It’s your life.
Live it Beautifully
visit us on the web - www.sonobello.com<https://outlook.office.com/owa/www.sonobello.com>
follow us on twitter: twitter.com/sonobello<https://outlook.office.com/owa/www.twitter.com/sonobello>
Body Contour Centers, LLC is not a medical provider and is not engaged in the practice of medicine. All medical decisions are the sole responsibility of Aesthetic Physicians, PC and Sono Bello physicians.
“Confidentiality Notice: This e-mail message, including any attachments, is for the sole use of the intended recipient(s) and may contain confidential and privileged information. Any unauthorized review, use, disclosure or distribution is prohibited. If you are not the intended recipient, please contact the sender by reply email or phone and destroy all copies of the original message.”
P Please consider the environment before printing this e-mail
Installed the correct technology probe - GE TruSignal
POST Test; Passed
Display picture quality; verified
Touch Screen; Calibrated
Updated time and date
Device information; Version and Network Settings; Verified
B1X5-REC recorder (S/N: _____VM23011084SA_____)
• Waveform 1: I
• Waveform 2: II
• Waveform 3: Resp
• The header line contains the date, time and some other applicable status; Verified
• The grid is clear; Verified
• The waveforms labels appear in the printout as configured; Verified
ECG Measurements:
80 bpm (± 5 bpm) = __80___ bpm
Pacemaker detection; Verified
Asystole detection; Verified
Leads off detection; Verified
Respiration Measurements:
20 RR (±5) = __20___ RR
Apnea Detection; Verified
SpO2 Measurements:
90% (±2%) = __90___%
NIBP Measurements:
200 mmHg (±1 mmHg) = ___200__mmHg
IBP Measurements:
0 mmHg = __0___ mmHg
200 mmHg (±10 mmHg) = ____200_ mmHg
Temperature Measurement:
98.6°F ± 0.18 °F = __98.6___ °F
Check Procedure Completed and Patient Discharged
PM Completed
🔍Technician Findings
Found that the Spo2 probe they is defective and the 2nd probe is the wrong technology need new spo2 probe
⚠️Problem Description
Hey there,
One of our vital sign machines won't read O2- we have swapped out several cords and it won't read. Can we get a loaner and send this out for repair? Thanks.
Asset ID#: 13367703
SN: SR723010125SA
GE B125M
https://www.ebiotrack.com/workorder.php?wo=69682
📄Description
Vital Signs Monitor, GE B40
SONO BELLOW CONSIGNED
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 11, 2025
To: emcentee@sonobello.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315129Location: OR-PortlandAddress: 5100 S Macadam Ave, Suite 180 Portland, OR 97239Device: Problem Description: Hey there,
One of our vital sign machines won't read O2- we have swa
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 4, 2025
To: emcentee@sonobello.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315129Location: OR-PortlandAddress: 5100 S Macadam Ave, Suite 180 Portland, OR 97239Device: Problem Description: Hey there,
One of our vital sign machines won't read O2- we have swa
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 27, 2025
To: emcentee@sonobello.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315129Location: OR-PortlandAddress: 5100 S Macadam Ave, Suite 180 Portland, OR 97239Device: Problem Description: Hey there,
One of our vital sign machines won't read O2- we have swa
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 20, 2025
To: emcentee@sonobello.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315129Location: OR-PortlandAddress: 5100 S Macadam Ave, Suite 180 Portland, OR 97239Device: Problem Description: Hey there,
One of our vital sign machines won't read O2- we have swa
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 13, 2025
To: emcentee@sonobello.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315129Location: OR-PortlandAddress: 5100 S Macadam Ave, Suite 180 Portland, OR 97239Device: Problem Description: Hey there,